This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
| Test | Result | No. | Positive rate of serological test (%) |
|||
|---|---|---|---|---|---|---|
| COV2T | sCOVG | ACOV2 | ACOV2S | |||
| SARS-CoV-2 RT-PCR | Positive | 115 | 87.0 | 79.1 | 43.5 | 86.1 |
| Negative | 71 | 90.1 | 85.9 | 46.5 | 94.4 | |
| Total | 186 | 88.2 | 81.7 | 44.6 | 89.2 | |
| Test | Sample | Repeatability |
Within-laboratory precision |
||||||
|---|---|---|---|---|---|---|---|---|---|
| Low |
High |
Low |
High |
||||||
| Mean±SD (U/mL) | CV (%) | Mean±SD (U/mL) | CV (%) | Mean±SD (U/mL) | CV (%) | Mean±SD (U/mL) | CV (%) | ||
| COV2T | QC material | <0.6 | - |
3.236±0.067 | 2.1 | <0.6 | - |
3.236±0.067 | 2.1 |
| sCOVG | QC material | 0.019±0.018 | 92.2 | 3.748±0.067 | 1.8 | 0.019±0.018 | 92.2 | 3.748±0.069 | 1.8 |
| ACOV2 | QC material | 0.089±0.002 |
2.4 | 3.518±0.036 | 1.0 | 0.089±0.003 | 2.9 | 3.518±0.063 | 1.8 |
| ACOV2S | QC material | <0.4 | - |
7.275±0.092 | 1.3 | <0.4 | - |
7.275±0.181 | 2.5 |
| Method x | Method y | Slope | Intercept (U/mL) | r |
|---|---|---|---|---|
| COV2T | sCOVG | 0.041 | 1.795 | 0.496 |
| sCOVG | ACOV2S | 142.7 | –33.840 | 0.833 |
| ACOV2S | COV2T | 6.352 | –4.449 | 0.413 |
| Comparing group | Concordance rate (%) | Agreement (κ) | p-value |
|---|---|---|---|
| COV2T vs. sCOVG | 93.5 | 0.750 | <0.001 |
| COV2T vs. ACOV2 | 51.5 | 0.096 | 0.028 |
| COV2T vs. ACOV2S | 96.8 | 0.839 | <0.001 |
| sCOVG vs. ACOV2 | 53.2 | 0.124 | 0.018 |
| sCOVG vs. ACOV2S | 92.5 | 0.700 | <0.001 |
| ACOV2 vs. ACOV2S | 51.1 | 0.098 | 0.019 |
SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; RT-PCR, reverse transcription polymerase chain reaction; COV2T, Atellica IM SARS-CoV-2 Total; sCOVG, Atellica IM SARS-CoV-2 IgG; ACOV2, Elecsys Anti-SARS-CoV-2; ACOV2S, Elecsys Anti-SARS-CoV-2 S.
SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; SD, standard deviation; CV, coefficient of variation; COV2T, Atellica IM SARS-CoV-2 Total; sCOVG, Atellica IM SARS-CoV-2 IgG; ACOV2, Elecsys Anti-SARS-CoV-2; ACOV2S, Elecsys Anti-SARS-CoV-2 S; QC, quality control. SDs and CVs could not be determined because the low QC material in the COV2T and ACOV2S had concentrations below the measuring range (<0.6 U/mL and <0.4 U/mL, respectively) throughout the experiment. The results of ACOV2 were presented as a cutoff index.
SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; COV2T, Atellica IM SARS-CoV-2 Total; sCOVG, Atellica IM SARS-CoV-2 IgG; ACOV2S, Elecsys Anti-SARS-CoV-2 S;
SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; COV2T, Atellica IM SARS-CoV-2 Total; sCOVG, Atellica IM SARS-CoV-2 IgG; ACOV2, Elecsys Anti-SARS-CoV-2; ACOV2S, Elecsys Anti-SARS-CoV-2 S.